Retsch AS 450 basic Manual de usuario Pagina 32

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 40
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 31
Cleaning and service
32
The AS450 basic is protected by a device fuse .
The external fusing must be T5.0A (230V) F15A (110V).
• Turn the device off at the mains (H).
• Pull the mains plug from the low heat device socket (J).
• Unscrew the fuse insert (I) and remove the fuse (IS).
CAUTION
A fuse must be replaced by one of the same type and with the same fuse value.
• Place the new fuse in the fuse insert.
NOTE
Do not tighten the fuse insert too much when screwing the fuse in!
• Screw the fuse insert (I) back in.
NOTE
If the fuse triggers again immediately after being replaced, there is a fault in the
device which must be rectified.
There are two fuses: one in the live conductor and one in the neutral conductor of
the power supply.
Pos: 14.1 /00010 Bedienungsa nleitungen Kapitels ammlungen/------- Seitenu mbruch ----------- @ 0\mod_1222344373 758_0.doc @ 23 86 @ @ 1
Vista de pagina 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 38 39 40

Comentarios a estos manuales

Sin comentarios